User Contributed 7411PC Procedure ============================================================================== INSTRUMENT: BALLANTINE 300: CALIBRATION /5100,5205 INSTRUMENT: BALLANTINE 300G: CALIBRATION /5100,5205 INSTRUMENT: BALLANTINE 300H: CALIBRATION /5100,5205 DATE: 15-Nov-88 AUTHOR: User Contributed REVISION: V 4.0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 35 NUMBER OF LINES: 102 CONFIGURATION: Fluke 5100B CONFIGURATION: Fluke 5205A ============================================================================== STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.004 JMPT 11.003 BALLANTINE 300G: CALIBRATION /5100,5205 1.006 JMPT 23.003 BALLANTINE 300H: CALIBRATION /5100,5205 1.007 HEAD {BALLANTINE MODEL 300 CALIB.} [7] 1.008 DISP Connect the "AC LINE CORD" to an AC POWER OUTLET and 1.008 DISP turn the "AC POWER SWITCH" on. Allow a 5 Min. warmup. 1.009 DISP Connect a standard DOUBLE BANANA PLUG to DOUBLE BANANA 1.009 DISP PLUG TEST CABLE to the UUT "INPUT TERMINALS". 1.010 HEAD {SCALE FACTOR ADJUSTMENT} [7] 1.011 ASK- F 1.012 MESS After SLEWING the STIMULUS for a UUT "METER" READING of 1.012 MESS exactly 10, select the "10V RANGE" and adjust "SCALE 1.012 MESS RANGE CONTROL" for a UUT "METER" READING of exactly 1. 1.012 MESS Return to the "1V RANGE" and repeat the procedure. 1.013 5100 1 1V 2% 500H 2W 2.001 MESS 2.002 HEAD {RANGE PERFORMANCE TEST} [7] 2.003 5100 0.01 0.01V 2% 500H 2W 3.001 5100 0.1 0.1V 2% 500H 2W 4.001 5100 1 1V 2% 50H 2W 5.001 5100 1 1V 2% 10KH 2W 6.001 5100 10 1V 2% 500H 2W 7.001 5100 10 2V 2% 500H 2W 8.001 5100 10 5V 2% 500H 2W 9.001 5100 10 10V 2% 500H 2W 10.001 5100 100 100V 2% 500H 2W 11.001 DISP This completes the MODEL 300 CALIBRATION. 11.002 JMP 35.001 11.003 HEAD {BALLANTINE MODEL 300G CALIB.} [7] 11.004 DISP Connect the "AC LINE CORD" to an AC POWER OUTLET and 11.004 DISP turn the "FUNCTION SELECTOR SWITCH" to "METER". 11.004 DISP Allow a 5 Min. warmup. 11.005 DISP Connect a TEST CABLE to the UUT "INPUT TERMINALS". 11.006 HEAD {SCALE FACTOR ADJUSTMENT} [7] 11.007 ASK- F 11.008 MESS Adjust "FULL SCALE ADJ. POT." on LEFT side of UUT 11.008 MESS "CASE" for a UUT "METER" READING of 10. Select the 11.008 MESS "10V RANGE" on the UUT and adjust the "MIN. SCALE ADJ. 11.008 MESS POT." on the RIGHT side of the UUT "CASE" for a UUT 11.008 MESS "METER" READING of 1. Reselect the "1V RANGE". Repeat. 11.009 5100 1 1V 1% 500H 2W 12.001 MESS 12.002 HEAD {RANGE PERFORMANCE TEST} [7] 12.003 5100 0.01 0.01V 1% 500H 2W 13.001 5100 0.1 0.1V 1% 500H 2W 14.001 5100 1 1V 1% 50H 2W 15.001 5100 1 1V 1% 500H 2W 16.001 5100 1 1V 1% 10KH 2W 17.001 5100 10 1V 1% 500H 2W 18.001 5100 10 2V 1% 500H 2W 19.001 5100 10 5V 1% 500H 2W 20.001 5100 10 10V 1% 500H 2W 21.001 5100 100 100V 1% 500H 2W 22.001 5205 1000 1000V 1% 500H 2W 23.001 DISP This completes the MODEL 300G CALIBRATION. 23.002 JMP 35.001 23.003 HEAD {BALLANTINE 300H CALIB.} [7] 23.004 DISP Connect the "AC LINE CORD" to an AC POWER OUTLET and 23.004 DISP turn the "FUNCTION SELECTOR SWITCH" to "METER". Allow 23.004 DISP a 5 Min. warmup. 23.005 DISP Connect a TEST CABLE to the UUT "INPUT TERMINALS". 23.006 HEAD {SCALE FACTOR ADJUSTMENT}[7] 23.007 ASK- F 23.008 MESS Adjust the "FULL SCALE ADJ. POT." on the LEFT side of 23.008 MESS the UUT "CASE" for a UUT "METER" READING of 30. Select 23.008 MESS the "30V RANGE" and adjust the "MIN. SCALE ADJ. POT" on 23.008 MESS the RIGHT side of the UUT "CASE" for a UUT "METER" 23.008 MESS READING of 3. Reselect the "3V RANGE" and repeat. 23.009 5100 3 3V 1% 500H 2W 24.001 MESS 24.002 HEAD {RANGE PERFORMANCE TEST} [7] 24.003 5100 0.003 0.003V 1% 500H 2W 25.001 5100 0.03 0.03V 1% 500H 2W 26.001 5100 0.3 0.3V 1% 500H 2W 27.001 5100 3 3V 1% 50H 2W 28.001 5100 3 3V 1% 500H 2W 29.001 5100 3 3V 1% 10KH 2W 30.001 5100 30 3V 1% 500H 2W 31.001 5100 30 6V 1% 500H 2W 32.001 5100 30 15V 1% 500H 2W 33.001 5100 30 30V 1% 500H 2W 34.001 5100 300 300V 1% 500H 2W 35.001 DISP This completes the BALLANTINE AC VTVM CALIBRATION. 35.002 END